Availability and performance of image/video-based vital signs monitoring methods: a systematic review protocol

Systematic Reviews20176:217

https://doi.org/10.1186/s13643-017-0615-3

Received: 1 June 2017

Accepted: 19 October 2017

Published: 25 October 2017

Abstract

Background

For many vital signs, monitoring methods require contact with the patient and/or are invasive in nature. There is increasing interest in developing still and video image-guided monitoring methods that are non-contact and non-invasive. We will undertake a systematic review of still and video image-based monitoring methods.

Methods

We will perform searches in multiple databases which include MEDLINE, Embase, CINAHL, Cochrane library, IEEE Xplore and ACM Digital Library. We will use OpenGrey and Google searches to access unpublished or commercial data. We will not use language or publication date restrictions. The primary goal is to summarise current image-based vital signs monitoring methods, limited to heart rate, respiratory rate, oxygen saturations and blood pressure. Of particular interest will be the effectiveness of image-based methods compared to reference devices. Other outcomes of interest include the quality of the method comparison studies with respect to published reporting guidelines, any limitations of non-contact non-invasive technology and application in different populations.

Discussion

To the best of our knowledge, this is the first systematic review of image-based non-contact methods of vital signs monitoring. Synthesis of currently available technology will facilitate future research in this highly topical area.

Systematic review registration

PROSPERO CRD42016029167

Keywords

MonitorNon-invasiveNon-contactImageCameraVital signs

Background

Monitoring of the basic functions of the human body is routine practice in healthcare settings from primary care to critical care. The term ‘vital signs’ typically refers to measurements including heart rate (HR), respiratory rate (RR), blood pressure (BP), oxygen saturations (SpO2) and temperature. Vital signs monitoring methods such as the electrocardiograph (ECG) and pulse oximetry have the advantages of being easily available, quick and non-invasive. In many clinical settings, these widely available monitors may be sufficient. However, they do have disadvantages which may limit their use in high dependency areas.

Examples of their limitations include the following:
  • ECG monitoring relies on application of three or more ECG electrodes on the patient. These consist of conducting gel surrounded by an adhesive area. Signal acquisition relies on good contact between the gel and the patient’s skin and this process is prone to errors from electrostatic and electromagnetic interference and skin impedance. The small amplitude bioelectrical signals measured are prone to interference from other biological signals or machinery.

  • A single point of measurement may limit the amount of information being collected [1]. A broader receiving system may offer more detailed spatial information simultaneously from multiple sites, allowing derivation/mapping of physiological parameters. Ultimately, this may facilitate insights that would be difficult to appreciate from a single-point analysis.

  • There are infection risks from being physically attached to monitors and the use of the same machines on multiple patients.

  • Monitors that require contact with the skin may lead to skin barrier breakdown. This is a concern in critical care patients who are more prone to hospital-acquired pressure ulcers and impaired skin healing [2], especially in neonates [3].

  • Multiple lines for various forms of monitoring create a hazard for both patients and staff, particularly during transfer. They may also interfere with clinical care.

  • Multiple contact monitors can interfere with sleep which is known to adversely affect recovery [4].

  • Non-contact monitoring is likely to be less emotionally traumatic for visitors in critical care settings [5].

Some of these limitations and complications may be avoided by using non-contact, non-invasive monitoring methods in the form of image/video-based monitoring. There is a huge potential for this type of monitoring in the current era of telemedicine, and the required technology has developed rapidly in the last decade. Several image-based technologies have been explored for vital signs monitoring, and it is our aim to provide a summary of their current availability and performance.

Objective

We aim to review non-contact, non-invasive image/video-based vital signs monitoring devices in health care and health-related fields. To our knowledge, this will be the first systematic review of the evidence for this type of monitoring. The review will provide a basis for further research using similar technology in a rapidly evolving field.

Methods/design

Protocol and registration

Methods of the systematic review have been developed in accordance with the Preferred Reporting Items for Systematic Reviews and Meta-Analyses (PRISMA) [6] guideline. The completed PRISMA-P [7] checklist is available as a supplementary file to this protocol (Additional file 1).

This proposed systematic review has been registered with PROSPERO (the International prospective register of systematic reviews): CRD42016029167.

Eligibility criteria

All papers looking at monitoring of HR, BP, RR or SpO2 using image analysis with comparison to a clinically validated reference device will be included without limitation on the setting. We will exclude temperature monitoring which lends itself to infrared imaging and has been extensively studied [810]. Technology which require placement of trackers or monitors on the body will be excluded. Studies using projected markers but require image/video analysis for monitoring purposes will be included. No limitations will be set on the primary objective of the studies; for example, studies aiming to refine image analysis will be included as long as they measure one of the specified vital signs and compare it to a clinically validated reference device as part of the methodology.

Initial screening process will select any literature reviews to ensure all original studies are included. Only human subject studies will be included, and no age limits will be set. Date and language restrictions will not be applied. Authors of studies published in languages other than English will be contacted by email for assistance with data extraction.

All unpublished studies found and articles in the media will be included wherever possible to minimise publication bias.

We will include randomised controlled trials, interventional studies, observational studies (including case-control or cohort studies) and pilot studies that compare an image-based monitoring system to any established forms of monitoring already used in clinical practice. Only studies that compare results from an image methodology with a clinically validated reference non-image methodology will be included. Any case series or reports will be included in the first instance with no limitations on the total number of subjects.

Outcomes

The primary intended outcome is a summary of the current availability of image/video-based vital signs monitoring methods and a comparison of the performance of the new technology with other clinically available methods.

Secondary outcomes of interest are the quality of the studies’ methodology with respect to the Guidelines for Reporting Reliability and Agreement Studies (GRRAS) [11], current limitations in image/video-based monitoring technology and the feasibility of its use in different populations.

It is probable that there will be considerable variability in the outcome measurements and the reporting of the technology’s efficacy. We will aim to produce a qualitative synthesis if the heterogeneity between the studies does not allow a quantitative summary.

Data sources/management

Searches will be performed in multiple databases including MEDLINE, Embase, CINAHL and Cochrane library. Given the subject topic, IEEE (Institute of Electrical and Electronics Engineers) Xplore Digital Library and ACM (Association for Computing Machinery) Digital Library will also be searched. OpenGrey will be searched for any unpublished grey literature. Given the popular culture/commercial applications of the technology, Google/Google Scholar will also be searched.

If any websites are available, primary original research data from which the technology is based will be sought.

Identified references will be downloaded to reference library software for the initial title and abstract screening. All references proceeding to the next stage will be handled using Distiller SR software (Evidence Partners, Ottawa, Canada). All searches will be saved for referencing, including a full list of papers and timeline of searches. The process of each paper will be clearly documented from the screening and review process.

Search strategy

The search strategy will be guided by a medical librarian who will advise on the conduct of the searches and limitations of the search terms used. Broad search terms will be used in order to capture all publications on this topic.

Initial search terms of interest include Monitor/Invasive/Non-invasive/Contact/Non-contact/Contactless/Remote/Image/Camera/Webcam/Web-cam/Physiologic/Measure/Respiratory rate/Breathing rate/Pulse/Heart rate/Blood pressure/SpO2.

Where appropriate, the terms will be exploded and MeSH (Medical Subject Headings) terms will be used. As demonstrated in the example, word stems will be used to generate results where appropriate.

Once searches are complete, the findings will be pooled and any duplicates removed.

Draft search strategy is outlined in Table 1.
Table 1

Example of search strategy

1

(remote*).ab,ti.

2

(measur*).ab,ti.

3

(imag*).ab,ti.

4

(monitor*).ab,ti.

5

(camera* OR webcam* OR web-cam* OR video).ab,ti.

6

(invasive* OR non-invasive* OR noninvasive OR contact* OR non-contact* OR noncontact* OR remote*).ab,ti.

7

(respira* OR respiratory rate OR breathing rate OR pulse OR heart rate OR blood pressure OR SpO2).ab,ti.

8

1 AND 2 AND 5

9

2 AND 3 AND 6 AND 7

10

4 AND 6 AND 7

11

5 AND 6

12

“Monitoring, Physiologic/methods”[MAJR]

13

5 AND 12

14

6 AND 12

15

8 OR 9 OR 10 OR 11 OR 13 OR 14

*Word truncation search

Study selection and data extraction

A four-stage review process will be used to select final articles for inclusion.
  1. Stage 1:

    Once searches are generated and checked for duplicates, the titles will be screened by two reviewers (MH, JC) to reject any which are clearly unrelated. Any articles with uncertainty or where there is a disagreement between the reviewers will be included for abstract review.

     
  2. Stage 2:

    Abstract review will be undertaken by two reviewers (MH, JC). Where there is a discrepancy, the article will proceed to a full-text review along with others selected by both reviewers.

     
  3. Stage 3:

    A selection of full-text articles will be assessed by two reviewers (MH, JC) using a pilot data extraction form on the reference management software. The extraction output will be compared and a final data extraction form will be created. All full texts will then be assessed by the two reviewers. Authors will be contacted by email for clarification or for further data wherever necessary.

     
  4. Stage 4:

    Once articles are selected, hand-searching strategy will be used to review their references to look for relevant articles not already identified. This is of particular importance as pseudo-healthcare use of non-contact monitoring or non-published articles may be identified using this method.

     
The returned papers will be categorised under the predominant technology used. Anticipated headings include the following:
  • Non-contact photoplethysmography (PPG)/imaging PPG

  • Thermal/infrared

  • Multi-slice image and tracking

  • Tracking of projected light

The above categories will be included in the data extraction form and effort will be made to group similar technologies together for analysis. We expect to stratify studies by predominant imaging modality, vital signs of interest and the target population.

Outcomes and prioritisation

The following data will be extracted from each study: the type and date of publication, funding source, conflict of interest, study type, vital sign of interest, setting, the number of participants, population and cohort data, eligibility criteria, image technology used, reference method, distance from camera to subject, body part imaged, frames per second, simulated physiological changes, performance of new technology compared to reference method, limitations reported, statistical analysis method, main conclusions and data to allow quality of reporting assessment according to GRRAS guidelines.

The primary intended outcome is a qualitative review of the availability and performance of different technology used for each vital sign. The studies will be categorised according to the video/image-based monitoring used (e.g. PPG, thermal) and the vital sign being measured (i.e. HR, BP, RR, SpO2) during the data extraction phase.

A secondary intended outcome is an appraisal of the quality of reporting among the method comparison studies included in the review with respect to the modified GRASS scoring system (Table 2). Furthermore, a summary of the current limitations of non-contact, non-invasive technology, potential methods of improving image acquisition and the application of the technology in different clinical populations will be synthesised.
Table 2

Adapted GRRAS tool for quality assessment of measurement comparison studies

Section

Original guidelines from GRRAS

Scoring

Title and abstract

Identify in title or abstract that interrater/intrarater reliability or agreement was investigated

N/Aa

 

Introduction

Name and describe the diagnostic or measurement device of interest explicitly

The image-based vital signs monitor is described

+

No description

Specify the subject population of interest

Specifies the subject population of interest

+

No population specified

Specify the rater population of interest (if applicable)

N/A

 

Describe what is already known about reliability/agreement and provide a rationale for the study

Describes what is already known about the reliability of image-based monitoring method and provide a rationale for the study

+

No statement on the current knowledge of the method and no rationale stated

Methods

Explain how the sample size was chosen. State the determined number of raters/subjects/objects/replicate observations

Explains how the sample size was chosen and/or state the determined number of subjects/replicate observationsb

+

No explanation of sample size. Number of subjects/replicate observations not stated

Describe sampling method

Describes sampling method

+

No description of sampling method

Describe the measurement/rating process (e.g. time interval between repeated measurements, availability of clinical information, blinding)

Describes the measurement process

+

No description of measurement process

State whether measurements/ratings were conducted independently

Two (or more) methods of measurements conducted independently

+

Measurements not conducted independently

Describe the statistical analysis

Describes the statistical analysis planned

+

No description of statistical analysis planned

Results

State the actual number of raters and subjects/objects which were included and the number of replicate observations which were conducted

States the actual number of subjects who were included and number of observations (e.g. duration of recording)

+

No statement of the number of subjects/observations

Describe the sample characteristics of raters/subjects

N/A

 

Report estimates of reliability and agreement including measure of statistical uncertainty

Reports estimates of reliability and agreement including measure of statistical uncertainty using gold standard measures

+

Reports estimates of reliability and agreement including measure of statistical uncertainty using acceptable standard measures

?

Reports estimates of reliability and agreement including measure of statistical uncertainty but using inappropriate measures/no estimates of reliability/agreement given

Discussion

Discuss the practical relevance of results

Discusses the practical relevance of results

+

No discussion of practical relevance of results

Auxillary material

Provide detailed results if possible (e.g. online)

N/A

 

N/A not applicable

aStudies will not be penalised for not stating the gold standard/reference method of monitoring in the title/abstract as we aim to include studies where this comparison was performed with an alternative primary aim (e.g. improve image analysis protocol)

bPilot studies will not be penalised for not stating how the sample size was chosen

Assessment of study quality

The appraisal will be performed by two reviewers (MH, JC) who will agree on the final assessment. In the event of a discrepancy, this will be resolved by discussion with a third reviewer (PW or DY).

Many studies of interest for the purpose of this review are likely to be experimental/pilot/proof-of-concept studies using a small number of subjects from a single centre. It is anticipated that it may not be possible to weight studies numerically in the final analysis due to the likely variability in the study designs and variable study sizes.

GRRAS recommendation for method comparison studies reporting is designed to cover a broad range of clinical test scores/classification/diagnosis and not all applicable for our use. Therefore, they have been adapted to suit our objectives (Table 2).

A key aspect of quality assessment is the appropriateness of the statistical analysis. Pearson’s correlation coefficient is not an appropriate statistical method to assess reliability and agreement, as originally discussed by Bland and Altman [12]. Despite this, a review of statistical methods used to test for agreement of medical instruments showed that the correlation coefficient remains the second most frequently used statistic [13]. In order to assess this aspect of each study, the statistical analysis will be checked against the published method agreement analysis guidelines [14]. The classification strategy based on the guidelines is summarised in Table 3. Each study will be assessed for statistical appropriateness. This has been incorporated to the modified GRRAS scoring system (Table 2).
Table 3

Classification of statistical methods used to compare image-guided vital signs measurement against reference device

Gold standard

Acceptable

Inappropriate

 • Bland-Altman plot/limits of agreement (LOA) analysis

 • Intraclass correlation coefficient

 • Lin’s concordance correlation coefficient

 • British standards reproducibility/repeatability coefficient

For all methods in this category, if repeat measurements per patient have not obviously been accounted for, then classify as acceptable.

• Mean square error/deviation

• Root mean square error/deviation

• Mean absolute error/deviation

• Accuracy

• Categorisation (e.g. proportion of time/measurements test within a threshold of gold standard)

• Descriptive only—plot of test against gold standard, or test and gold standard signals overlaid

• (Pearson’s) correlation coefficient

T test

Where one method is used, or all methods used are in the same category, the overall rating will be given as that category. Where multiple methods are used and they are in adjacent categories, the study will be categorised as the better. If gold standard and inappropriate methods are used together, the study will be classified as acceptable.

It is expected that Bland and Altman’s method will be used most frequently for measuring agreement. Specifically, it is important that where multiple observations are made per individual (e.g. multiple HR measurements from one subject using two methods), this is accounted for in the statistical assessment [15]. This has been incorporated into the statistical rating scale (Table 3).

Each included study will be given a rating based on the modified GRASS scoring system as below:
  • All items are “+”: high quality

  • All items are “+” except the statistical analysis rating of “?” OR one of the other sections is given “-”: moderate quality

  • Statistical analysis is given “-” OR two or more sections are given “-”: low quality

Finally, the GRADE methodology will be used to systematically rate the strength of the overall review as high/moderate/low/very low [16].

Heterogeneity

There is likely to be a high degree of heterogeneity across the studies given the likely variation in reported outcomes. It is hoped that it would be possible to compare methods post stratification by the vital signs of interest (e.g. HR) and by the predominant technology used.

Missing data

We will seek missing data by contacting original authors. Effort will be made to include all available studies.

Subgroup analysis

Subgroup analysis will be undertaken for the group of studies with acceptable statistical methods and outcomes summarised in numerical format if possible. Should the extracted data be deemed sufficiently homogenous from a clinical standpoint for quantitative synthesis to be undertaken, we anticipate performing numerical analysis. If we identify studies that are sufficiently homogenous, we will attempt a meta-analysis of the results, using random-effects methods to combine measures of bias and agreement from multiple Bland-Altman analyses [17]. Similar techniques may also be used to combine measures of accuracy and error. If significant heterogeneity exists between the types of study making the numerical subgroup analysis impossible, the analysis will be qualitative. If there is sufficient stratified data for the application or performance of any particular imaging modality or detected vital sign, a statistical subgroup analysis will be performed.

Amendments to protocol

Any deviation from this protocol will be dated and documented to ensure a transparent review process. As per PRISMA guidelines, no changes will be made to the main body of the protocol and any unanticipated additional findings will be discussed in the final review.

Discussion

This systematic review will synthesise a summary of the current technology available for image-based monitoring of vital signs.

The main strength of the proposed review is that a summary of the availability and effectiveness of non-invasive, non-contact monitoring methods and their utility in the clinical environment will provide a basis for the direction of future work in the field. Furthermore, an appraisal of the quality of reporting among these studies will guide authors for reporting of future studies. Similarly, a review of the statistical methods employed among these studies will improve the analysis of future studies.

Limitations

Despite the outlined wide search strategy, we anticipate a number of difficulties in carrying out the proposed systematic review. There is high likelihood of publication bias as the published studies/imaging methods are more likely to have better performance compared to those that are not published. We also anticipate that studies conducted as part of doctoral theses may not be found despite the wide search strategy and/or may be difficult to access. It is likely that a proportion of these studies will have been published in peer-reviewed journals with further likelihood of publication bias. Every effort will be made to include grey literature available with review of citations/references and advice from a librarian.

We also anticipate that a large number of studies included in the review will be proof-of-concept studies with very few subjects. Therefore, we anticipate a heterogeneous group of studies to be included in the review which may make quantitative summary difficult.

We believe that this review will provide a useful summary of the current technologies available for image-based monitoring of vital signs and guide the direction of future research for more complex monitoring, especially in critical care settings.

Abbreviations

BP: 

Blood pressure

ECG: 

Electrocardiograph

GRRAS: 

Guidelines for Reporting Reliability and Agreement Studies

HR: 

Heart rate

ICU: 

Intensive care unit

LOA: 

Limits of agreement

PPG: 

Photoplethysmography

PRISMA: 

Preferred Reporting Items for Systematic Reviews and Meta-Analyses

RR: 

Respiratory rate

SpO2

Oxygen saturations

Declarations

Acknowledgements

The authors would like to thank Tatjana Petrinic for her guidance during development of search strategy.

Funding

This work is supported by the National Institute for Health Research. The funder supports salaries for this work but has had no role in the development of this protocol.

Availability of data and materials

Not applicable

Authors’ contributions

MH and PW conceived the original idea. MH and JC developed the search strategy. MH and SG developed the statistical assessment and analysis strategy. MH has written this manuscript with revisions by JC, SG, DY and PW. All authors read and approved the final manuscript.

Ethics approval and consent to participate

Not applicable

Consent for publication

Not applicable

Competing interests

The authors declare that they have no competing interests.

Publisher’s Note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.

Open AccessThis article is distributed under the terms of the Creative Commons Attribution 4.0 International License (http://creativecommons.org/licenses/by/4.0/), which permits unrestricted use, distribution, and reproduction in any medium, provided you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons license, and indicate if changes were made. The Creative Commons Public Domain Dedication waiver (http://creativecommons.org/publicdomain/zero/1.0/) applies to the data made available in this article, unless otherwise stated.

Authors’ Affiliations

(1)
Kadoorie Centre for Critical Care and Trauma Research and Education, John Radcliffe Hospital
(2)
Nuffield Department of Clinical Neurosciences, University of Oxford
(3)
Centre for Statistics in Medicine, University of Oxford

References

  1. Sun Y, Papin C, Azorin-Peris V, Kalawsky R, Greenwald S, Hu S. Use of ambient light in remote photoplethysmographic systems: comparison between a high-performance camera and a low-cost webcam. J Biomed Opt. 2012;17(3):037005.View ArticlePubMedGoogle Scholar
  2. Mestha LK, Kyal S, Xu B, Lewis LE, Kumar V. Towards continuous monitoring of pulse rate in neonatal intensive care unit with a webcam. Conf Proc IEEE Eng Med Biol Soc. 2014:3817–20.Google Scholar
  3. Johnson DE. Extremely preterm infant skin care: a transformation of practice aimed to prevent harm. Adv Neonatal Care. 2016;16(Suppl 5S):S26–32.View ArticlePubMedGoogle Scholar
  4. Kamdar BB, Needham DM, Collop NA. Sleep deprivation in critical illness: its role in physical and psychological recovery. J Intensive Care Med. 2012;27(2):97–111.View ArticlePubMedGoogle Scholar
  5. Hardicre J. Meeting the needs of families of patients in intensive care units. Nurs Times 2003;99(27):26-27.Google Scholar
  6. Moher D, Liberati A, Tetzlaff J, Altman DG, The PRISMA Group. Preferred Reporting Items for Systematic Reviews and Meta-Analyses: the PRISMA Statement. PLoS Med. 2009;6(7):e1000097. doi: 10.1371/journal.pmed1000097.View ArticlePubMedPubMed CentralGoogle Scholar
  7. Moher D, Shamseer L, Clarke M, Ghersi D, Liberati A, Petticrew M, Shekelle P, Stewart LA. Preferred Reporting Items for Systematic Review and Meta-Analysis Protocols (PRISMA-P) 2015 Statement. Syst Rev. 2015;4(1):1. doi: 10.1186/2046-4053-4-1.View ArticlePubMedPubMed CentralGoogle Scholar
  8. Aubakir B, Nurimbetov B, Tursynbek I, Varol HA. Vital sign monitoring using Eulerian video magnification and thermography. Conf Proc IEEE Eng Med Biol Soc. 2016:3527–30.Google Scholar
  9. Wang K, Gill P, Wolstenholme J, Price CP, Heneghan C, Thompson M, Pluddemann A. Non-contact infrared thermometers for measuring temperature in children: primary care diagnostic technology update. Br J Gen Pract. 2014;64(627):681–3.View ArticleGoogle Scholar
  10. Lam H, Kostov Y, Tolosa L, Falk S, Rao G. High resolution non-contact fluorescence based temperature sensor for neonatal care. Meas Sci Technol. 2012;23(3):035104.View ArticlePubMedPubMed CentralGoogle Scholar
  11. Kottner J, Audige L, Brorson S, Donner A, Gajewski BJ, Hrobjartsson A, Roberts C, Shoukri M, Streiner DL. Guidelines for Reporting Reliability and Agreement Studies (GRRAS) were proposed. J Clin Epid. 2011;64:96–106.View ArticleGoogle Scholar
  12. Bland JM, Altman DG. Statistical methods for assessing agreement between two methods of clinical measurement. Lancet. 1986;1(8476):307–10.View ArticlePubMedGoogle Scholar
  13. Zaki R, Bulgiba A, Ismail R, Ismail NA. Statistical methods used to test for agreement of medical instruments measuring continuous variables in method comparison studies: a systematic review. PLoS One. 2012;7(5):e37908.View ArticlePubMedPubMed CentralGoogle Scholar
  14. Watson PF, Petrie A. Method agreement analysis: a review of correct methodology. Theriogenology. 2010;7:1167–79.View ArticleGoogle Scholar
  15. Bland JM, Altman DG. Agreement between methods of measurement with multiple observations per individual. J Biopharm Stat. 2007;17(4):571–82.View ArticlePubMedGoogle Scholar
  16. Belsham H, Helfand M, Schunemann HJ, Oxman AD, Kunz R, Brozek J. GRADE guidelines: 3. Rating the quality of evidence. J Clin Epidemiol. 2011;64:401–6.View ArticleGoogle Scholar
  17. Tipton E, Shuster J. A framework for the meta-analysis of Bland-Altman studies based on a limits of agreement approach. Stat Med. 2017; doi: 10.1002/sim.7352. [Epub ahead of print]

Copyright

© The Author(s). 2017